Compliance Verification

How to Ensure Your Ionizer is Working Correctly

The best way to keep electrostatic sensitive devices (ESDs) from damage is to ground all conductive objects and remove insulators from your ESD Protected Area (EPA). This is not always possible because some insulators are “process-essential” and are necessary to build or assemble the finished product. The only way to control charges on these necessary non-conductive items is the use of ionization systems.

However, if an ionizer is out of balance, instead of neutralizing charges, it will produce primarily positive or negative ions. This results in placing an electrostatic charge on items that are not grounded, potentially discharging and causing ESD damage to nearby sensitive items.

It is therefore essential to regularly clean your ionizers and verify they function correctly. Below we have put together a list of tasks you need to perform with your ionizers on a regular basis.

Maintenance

All ionization devices will require periodic maintenance for proper operation. Maintenance intervals for ionizers vary widely depending on the type of ionization equipment and use environment. Critical clean room uses will generally require more frequent attention. It is important to set-up a routine schedule for ionizer service. Routine service is typically required to meet quality audit requirements.” (ESD Handbook TR20.20 section 5.3.6.7 Maintenance / Cleaning)

EIA-625, recommends checking ionizers every 6 months, but this may not be suitable for many programs particularly since an out-of-balance may exist for months before it is checked again. ANSI/ESD S20.20 section 6.1.3.1 Compliance Verification Plan Requirement states: “Test equipment shall be selected to make measurements of appropriate properties of the technical requirements that are incorporated into the ESD program plan.

Under normal conditions, an ionizer will attract dirt and dust (especially on the emitter points). To maintain optimum neutralization efficiency and operation, cleaning should be performed on a regular basis.

1. Case

Wipe the case with a soft cloth and deionized water. Fully squeeze the wiping cloth or sponge to remove any excess liquid. If a stronger cleaning solution is required, dab a soft cloth with mixture of isopropyl alcohol and deionized water (70% IPA and 30% DI water).

2. Emitter Points

The emitter points should be cleaned using specific emitter point cleaners or a swab dampened with Isopropyl alcohol. Below are general instructions on how to clean emitter points. However, each unit is slightly different so always refer to the ionizer’s manual.

  1. Turn the unit OFF and unplug the power cord.
  2. Open the top screen by loosening the screw and swinging the grill to one side.
  3. Clean the emitter points using the an emitter point cleaner or a swab dampened with Isopropyl alcohol.
  4. Re-attach the top screen.
  5. Plug in the power cord and turn the unit ON.
  6. Verify the performance of the ionizer by using a charged plate monitor or ionization test kit (see below).

Cleaning of Emitter Points using SCS 9110-NO as an example

With normal handling, the emitter points should not require replacement during the life of the unit.

Check out how to clean an Ionizer in A Minute with Miranda:

Verification

Per ESD TR53 section 5.3.6.7.1 “The best practice is to measure the offset voltage and discharge times, clean the unit, including emitter points and air filters if present, offset voltage to zero (if adjustable), and then repeat offset voltage and discharge time testing. If the unit does not meet offset voltage specifications or minimum established discharge time limits, further service is indicated. Manufacturers should provide details on service procedures and typical service intervals.

Most companies will assign a number or otherwise identify each ionizer and setup a Compliance Verification / Maintenance / Calibration schedule. If the ionizers all test good, the data can justify lengthening the calibration period. If ionizers require adjustment, the calibration period should be shortened. Although ESD TR53 does not advise a test frequency, JESDD625-A (Revision of EIA-625) recommends ionizers be tested semi-annually, noting to use “S3.1 except the number of measurement points and locations may be selected based on the application.

Verification should be performed in accordance with the ESD Association ionization standard ANSI/ESD STM3.1.

Below are general instructions on how to verify your ionizer’s offset voltage and discharge time. Always refer to the User Guide accompanying your Charge Plate Monitor or Ionization Test Kit for proper operation and setup.

1. Testing Ionizer Offset Voltage:

The required limit per ANSI/ ESD S20.20 is less than ± 35 volts. Check your ionizer’s operating manual or consult with the ionizer manufacturer to determine what the offset voltage should be for your ionizer.

Charge Plate Monitor (CPM)

  1. Position the ionizer and charge plate monitor as shown below.
  2. Set the CPM to Decay/Offset mode.
  3. Set the CPM to decay and offset voltage mode with a starting charge at either + or – 1 KV and a stopping charge at either + or -100 Volts.
  4. Start the decay/offset test sequence on the CPM. This will take a few seconds.
  5. Record the decay time, and offset voltage as displayed on the CPM.

Positioning your Charge Plate Monitor for Overhead and Benchtop Ionizers

Ionization Test Kit

  1. Zero the charge plate by touching it with a grounded object. This can either be the finger of a grounded person or some other item which is connected to electrical ground. In either case, zeroing the charge plate should make the display on the field meter read zero.
  2. Hold the meter approximately one foot (30.5 cm) in front of the ionizer.
  3. Monitor the display. The value displayed is the offset balance of the ionizer, which is the difference between the number of positive and negative ions being emitted.
Testing Ionizer Offset Voltage using the SCS 718A Air Ionizer Test Kit

Portable Charge Plate Monitor

  1. Turn the meter on.
  2. Zero point adjust the meter. Bring the plate into contact with a grounded body, and press the 0ADJ button.
  3. Set the desired mode. Press the SELECT button, and the meter will scroll through the modes. To measure balance (offset voltage), use I.B mode.
  4. Point the plate at the measurement location, the meter will measure the balance.

Watch how to test the balance using the SCS Portable Charge Plate Monitor:

2. Testing Ionizer Discharge Time:

The required limit per ANSI/ESD S20.20 is “user defined”. Please refer to the ionizer’s operating manual or consult with the ionizer manufacturer to determine what this discharge time should be.

Charge Plate Monitor (CPM)

  1. Set the CPM to Decay/Offset mode.
  2. Set the CPM to decay and offset voltage mode with a starting charge at either + or – 1 KV and a stopping charge at either + or -100 Volts.
  3. Start the decay/offset test sequence on the CPM. This will take a few seconds.
  4. Record the decay time, and offset voltage as displayed on the CPM.

Ionization Test Kit

  1. After charging the plate of the ionization test kit, hold the field meter approximately one foot (30.5 cm) away from the ionizer.
  2. Monitor the display of the meter to see how quickly the 1.1 kV charge is dissipated to 0.1 kV.
  3. The speed at which this occurs (the discharge time) indicates how well the ionizer is operating.
  4. Repeat this procedure for both a positively and a negatively charged plate.

Portable Charge Plate Monitor

  1. If you haven’t already, turn the meter on and zero it.
  2. Set the desired decay range, select +Decay to measure positive discharge time. Select -Decay to measure negative discharge time.
  3. Point the plate in the airflow of the ionizer to be measured and press the START button.
  4. The meter will chirp and display the neutralization time when the measurement is completed.

Watch how to test the decay time using the SCS Portable Charge Plate Monitor:

Some ionizers offer adjustment options (e.g. trim pots) which allow modification of the offset voltage.

However, if your ionizer is out of balance (and cannot be adjusted) or if the discharge time is out of specification, the ionizer will require service/repair by an authorized company.

Conclusion

Ionization is one of the best methods of removing charges from insulators and as a result plays an important role in controlling ESD.

Remember though: ionizers require periodic cleaning of emitter pins and verifying of the offset voltage and discharge time. Otherwise, instead of neutralizing charges, the ionizer will primarily produce positive or negative ions. The ionizer will therefore place an electrostatic charge on items that are not grounded, potentially discharging and causing ESD damage to nearby sensitive items.

Measuring Effectiveness of an ESD Control Program

Introduction

Electronic devices and systems can be damaged by exposure to high electric fields as well as by direct electrostatic discharges. A good circuit layout and on-board protection may reduce the risk of damage by such events, but the only safe action at present is to ensure that devices are not exposed to levels of static electricity above the critical threshold.

This can only be achieved by introducing a static control program which usually involves setting up an ESD Protected Area (EPA) in which personnel are correctly grounded and all meet the ESD Standard. However, setting up an EPA does not of itself guarantee a low static environment. Production procedures may change, new materials may be introduced, the performance of older materials may degrade and so on.

Measuring Effectiveness of an ESD Control Program

To ensure the effectiveness of any static control program it is important that regular measurements are carried out:

  1. to determine the sensitivity to ESD of devices being produced or handled.
  2. to confirm that static levels are lower than the critical level, and that new or modified work practices have not introduced high static levels.
  3. to ensure that both new and existing materials in the EPA meet the necessary requirements.

Only after an ‘operational baseline’ has been established by regular auditing will it become possible to identify the origin of unexpected problems arising from the presence of static.

1. Determining the sensitivity of ESD sensitive Devices

It is important to understand the sensitivity of ESD sensitive devices before an action plan can be created. Once you know the sensitivity of the items you are handling, can you work towards ensuring you’re not exceeding those levels.

Part of every ESD control plan is to identify items in your company that are sensitive to ESD. At the same time, you need to recognize the level of their sensitivity. As explained by the ESD Association, how susceptible to ESD a product is depends on the item’s ability to either:

  • dissipate the discharge energy or
  • withstand the levels of current.

2. Measurements to prove the effectiveness of an ESD Control Program

Measuring electrostatic quantities poses special problems because electrostatic systems are generally characterized by high resistances and small amounts of electrical charge. Consequently, conventional electronic instrumentation cannot normally be used.

Measuring Electrical Field

Wherever electrostatic charges accumulate, they can be detected by the presence of an associated electric field. The magnitude of this field is determined by many factors, e. g. the magnitude and distribution of the charge, the geometry and location of grounded surfaces and the medium in which the charge is located.

The current general view of experts is that the main source of ESD risk may occur where ESDS can reach high induced voltage due to external fields from the clothing, and subsequently experience a field induced CDM type discharge.” [CLC TR 61340-5-2 User guide Garments clause 4.7.7.1 Introductory remarks]

718_Use2.jpg
Using the 718 Static Sensor to test static fields

A static field meter is often used for ESD testing of static fields. It indicates surface voltage and polarity on objects and is therefore an effective problem-solving tool used to identify items that are able to be charged.

A field meter can be used to:

  • verify that automated processes (like auto insertion, tape and reel, etc.) are not generating charges above acceptable limits.
  • measure charges generated by causing contact and separation with other materials.
  • demonstrate shielding by measuring a charged object and then covering the charged item with an ESD lab coat or shielding bag. Being shielded the measured charge should be greatly reduced.

 

Measuring ESD Events

ESD events can damage ESD sensitive items and can cause tool lock-ups, erratic behavior and parametric errors. An ESD Event Detector like the EM Eye ESD Event Meter will help detect most ESD events. It detects the magnitude of events and using filters built into the unit, it can provide approximate values for some ESD events for models (CDM, MM, HBM) using proprietary algorithms.

Using the EM Eye ESD Event Meter to detect ESD Events

Solving ESD problems requires data. A tool counting ESD events will help carry out a before-and-after analysis and will prove the effectiveness of implementing ESD control measures.

 

3. Checking Materials in your EPA

When talking about material properties, the measurement you will most frequently come across is “Surface Resistance”. It expresses the ability of a material to conduct electricity and is related to current and voltage. The surface resistance of a material is the ratio of the voltage and current that’s flowing between two pre-defined electrodes.
It is important to remember that the surface resistance of a material is dependent on the electrodes used (shape as well as distance). If your company implements an ESD control program compliant to the ESD Standard ANSI/ESD S20.20, it is therefore vital to carry out surface resistance measurements as described in the Standard itself. For more information on the definition of resistance measurements used in ESD control, check out this post.

A company’s compliance verification plan should include periodic checks of surfaces measuring:

  • Resistance Point-to-Point (Rp-p) and
  • Resistance-to-ground (Rg).

SRMeter2_use.jpg
Measuring Surface Resistance of worksurface matting using the
SRMETER2 Surface Resistance Meter

Surface resistance testers can be used to perform these tests in accordance with ANSI/ESD S20.20 and its test method ANSI/ESD S4.1; if these measurements are within acceptable ranges, the surface and its connections are good. For more information on checking your ESD control products, catch-up with this. It goes into depth as to what products you should be checking in your EPA and how they should be checked.

 

Conclusion

Measurements form an integral part of any ESD control program. Measuring devices help identify the sensitivity of ESD devices that ESD programs are based on, and also are used to verify the effectiveness of ESD control programs set in place. High quality instruments are available commercially for measuring all the parameters necessary for quantifying the extent of a static problem.

We hope the list above has introduced the techniques most commonly used. For more information on how to get your ESD control program off the ground, Request a free ESD/EOS Assessment at your facility by one of our knowledgeable local representatives to evaluate your ESD program and answer any ESD questions!

 

 

5 Common Mistakes in ESD Control & How to Avoid Them

Many companies implement an ElectroStatic Discharge (ESD) Control Program with the aim of improving their operations. Effective ESD control can be a key to improving:

  • Productivity
  • Quality
  • Customer Satisfaction

Problems arise when an organization invests in ESD protective products and/or equipment and then misuses them. Misuse of ESD protective products and/or equipment wastes invested money and can also be causing more harm than good. Today’s blog post will highlight some of the major issues we have come across and how you can avoid or fix them.

About ESD Control and ESD Protection

Remember that for a successful ESD control program, ESD protection is required throughout the manufacturing process: from goods-in to assembly all the way through to inspection. Anybody who handles electrical or electronic parts, assemblies or equipment that are susceptible to damage by electrostatic discharges should take necessary precautions.

Just like viruses or bacteria that can infect the human body, ESD can be a hidden threat unable to be detected by human eyes. Hidden viral/bacterial threats in hospitals are controlled by extensive contamination control procedures and protective measures such as sterilization. The same principles apply to ESD control: you should never handle, assemble or repair electronic assemblies without taking adequate protective measures against ESD.

Common Mistakes in ESD Control

1. Ionizers are poorly maintained or out-of-balance

If an ionizer is out of balance, instead of neutralizing charges, it will produce primarily positive or negative ions. This results in placing an electrostatic charge on items that are not grounded, potentially discharging and causing ESD damage to nearby sensitive items.

Step3 Remember to clean emitter pins and filters using appropriate tools. Create a regular maintenance schedule which will extend the lifespan of your ionizers tremendously.

Consider using ionizers with “Clean Me” and//or “Balance” alarms. These will alert you when maintenance is required.

Step2.png All ionization devices will require periodic maintenance for proper operation. Maintenance intervals for ionizers vary widely depending on the type of ionization equipment and use environment. Critical clean room use will generally require more frequent attention. It is important to set up a routine schedule for ionizer service.”

[ESD TR20.20 Handbook Ionization clause 15.8 Maintenance / Cleaning]

If you would like to learn more about how ionizers work and what type of ionizer will work best for your application, check out this post for detailed coverage.

2. ESD Garments are Ungrounded

We’ve seen it so many times: operators wearing an ESD coat (without appropriate wrist straps and/or footwear/flooring) thinking they are properly grounded. However, without proper electrical bonds to a grounding system they are not grounded!

Step3 Every ESD garment needs to be electrically bonded to the grounding system of the wearer. Otherwise it just acts as a floating conductor. There are a few options to choose from:

  • Wrist Straps
  • ESD footwear/flooring
  • Hip-to-Cuff grounding
Step2 After verifying that the garment has electrical conductivity through all panels, the garment should be electrically bonded to the grounding system of the wearer so as not to act as a floating conductor.

This can be accomplished by several means:

  1. Ground the garment to the body through a wrist strap-direct connection with an adapter.
  2. Ground the garment through conductive wrist or heel cuffs in direct contact with the skin of a grounded operator.
  3. Ground the garment through a typical separate ground cord, directly attached to an identified groundable point on the garment.
  4. Garments should be worn with the front properly snapped or buttoned to avoid exposure of possible charge on personal clothing worn under the garment.

[ESD TR20.20 Handbook Garments clause 19.4 Proper Use]

ESD clothing loses their ESD properties over time. It is therefore an important part of the ESD Control Program to incorporate periodic checks (see #3 below) of ESD garments.

If you need more information on ESD garments, we recommend having a look at this post.

3. No Compliance Verification Plan / Not Checking ESD Control Products

Companies can invest thousands of dollars in purchasing and installing ESD control products but then waste their investment by never checking their ESD items. This results in ESD equipment that is out of specification. Without the tools in place to check their ESD items, companies may have no idea if they are actually working correctly. Remember: ESD products (like any other product) are subject to wear and tear, and other errors when workstations get moved, ground cords get disconnected…etc. The list goes on.

Step3 When investing in ESD control products, make sure you also establish a Compliance Verification Plan. This ensures that:

  • ESD equipment is checked periodically
  • Necessary test equipment is available
Step2 A compliance verification plan shall be established to ensure the organization’s fulfilment of the requirements of the plan. Process monitoring (measurements) shall be conducted in accordance with a compliance verification plan that identifies the technical requirements to be verified, the measurement limits and the frequency at which those verifications shall occur. The compliance verification plan shall document the test methods used for process monitoring and measurements. If the organization uses different test methods to replace those of this standard, the organization shall be able to show that the results achieved correlate with the referenced standards. Where test methods are devised for testing items not covered in this standard, these shall be adequately documented including corresponding test limits. Compliance verification records shall be established and maintained to provide evidence of conformity to the technical requirements.
The test equipment selected shall be capable of making the measurements defined in the compliance verification plan.
”[ANSI/ESD clause 7.4 Compliance verification plan]

We provide detailed instructions on how to create a Compliance Verification Plan in this post.

4. Improperly Re-Using Shielding Bags / Using Shielding Bags with Holes or Scratches

ESD Shielding Bags are used to store and transport ESD sensitive items. When used properly, they create a Faraday Cage effect which causes charges to be conducted around the outside surface. Since similar charges repel, charges will rest on the exterior and ESD sensitive items on the inside will be ‘safe’. However, if the shielding layer of an ESD Shielding Bag is damaged, ESD sensitive items on the inside will not be protected anymore.

Step3 Re-using shielding bags is acceptable as long as there is no damage to the shielding layer. Shielding bags with holes, tears or excessive wrinkles should be discarded.

Use a system of labels to identify when the bag has gone through five (5) handling cycles. When there are five broken labels, the bag is discarded.

Step2 ESD shielding packaging is to be used particularly when transporting or storing ESD sensitive items outside an ESD Protected Area.

Transportation of ESDS items outside an ESD Protected Area (hereafter referred to as “EPA”) requires enclosure in static protective materials, although the type of material depends on the situation and destination. Inside an EPA, low charging and static dissipative materials may provide adequate protection. Outside an EPA, low charging and static discharge shielding materials are recommended. While these materials are not discussed in

the document, it is important to recognize the differences in their application. For more clarification see ANSI/ESD S541.

[ANSI/ESD Foreword]

This post provides further “dos and don’ts” when using ESD Shielding Bags.

5. Using Household Cleaners on ESD Matting

The use of standard household cleaners on ESD matting can put an ESD Control Program at risk and damage the ESD properties of items. Many household cleaners contain silicone or other insulative contaminants which create that lovely shine you get when wiping surfaces in your home. The problem is that silicone and other chemical contaminates can create an insulative layer which reduces the grounding performance of the mat.

Step3 Don’t spend all this extra money on ESD matting and then coat it with an insulative layer by using household cleaners. There are many specially formulated ESD surface and mat cleaners available on the market. Only clean your ESD working surfaces using those cleaners.
Step2 “Periodic cleaning, following the manufacturer’s recommendations, is required to maintain proper electrical function of all work surfaces. Ensure that the cleaning products used to not leave an electrically insulative residue which is common with some household cleaners that contain silicone.”

[ESD TR20.20 Handbook Worksurfaces clause 10.5 Maintenance]

Conclusion

There are many more issues we see when setting foot into EPAs and the above list is by no means complete. These are the most common issues we’ve found when assessing EPAs.

It is important to train all personnel using ESD products and/or equipment to follow proper ESD control programs, and maintenance procedures to avoid common ESD control mistakes. Basic ESD control principles should be followed for an ESD control program to be successful:

  • Ground conductors.
  • Remove, convert or neutralize insulators with ionizers.
  • Shield ESD sensitive items when stored or transported outside the EPA.

What mistakes do you commonly see when walking through an EPA? Let us know what you commonly see in the comments and your solutions for fixing them!

For more information on how to get your ESD control program off the ground and create an EPA, check this post.

6 Tips for handling “Class 0” Items

When talking about ESD Classifications a little while ago, we identified a “class 0” item as withstanding discharges of less than 250 volts.

The introduction of ANSI/ESD S20.20 states: “This standard covers the requirements necessary to design, establish, implement and maintain an Electrostatic Discharge (ESD) Control Program for activities that manufacture, process, assemble, install, package, label, service, test, inspect or otherwise handle electrical or electronic parts, assemblies and equipment susceptible to damage by electrostatic discharges greater than or equal to 100 volts Human Body Model (HBM) and 200 volts Charged Device Model (CDM).

So how do you handle items that are susceptible to voltages of less than 100V? That’s what we’re going to answer in today’s blog post.

 

Introduction

Years ago, it was common for devices to be vulnerable to voltages greater than 100 V. As the need for smaller and faster devices increased, so did their sensitivity to ElectroStatic Discharges as circuit-protection schemes were removed to stay ahead of the market. These new extremely sensitive components are now susceptible to discharges nearing 0 V. This causes problems for companies handling these devices: while their ESD program may be in compliance with the ESD Standard, extremely sensitive devices require tighter ESD Control to protect them from ESD failures.

 

What is a “Class 0” device?

Before moving any further, we need to qualify the term “class 0”. As stated above, the HBM Model refers to any item with a failure voltage of less than 250 V as a “class 0” component. However, in recent times, the term has been used more and more to describe ultra-sensitive devices with failure voltages of less than 100 V. Whilst the following tips and tricks work on any “class 0” item, they are specifically designed to protect extremely sensitive items that withstand discharges of less than 100 V.

esd_sterilization.jpg
Ultra-sensitive devices are extremely common

Before Updating Your ESD Program

“Class 0” refers to a wide range of items and there are a few things you should remember before making any changes to your existing ESD program:

  1. Verify what ESD Model your company/engineers/customers etc. are referring to. As we have learnt in the past, there are different ESD models (HBM, CDM, MM) as well as individual classifications for each model. A lot of people get confused when it comes to citing ESD classifications. There is only one “class 0” which refers to the human body model (HBM) but it’s always best to check.
  2. Check the specific withstand voltage an individual part is susceptible to. “Class 0” refers to all items that withstand discharges of less than 250 V. However, there is a big difference between a failure voltage of 240 V or 50 V. You need to have detailed ESD sensitivity information available before being able to make decisions on how to improve your existing ESD control program. This step is part of creating a compliance verification plan.
  3. A part’s ESD classification is only of importance until it is ‘merged’ into an assembly. So, the ESD classification of a device only refers to the stand-alone component. Once it goes into another construction, the classification of the whole assembly is likely to change.

 

Tips for handling “Class 0” Items

Below are 6 tips that will help your company to upgrade your ESD control program so you can effectively and efficiently handle ultra-sensitive items without risking ESD damage.

One thing to note: The best approach to stay ahead of the game is taking proactive actions. It is critical to figure out how to protect your components from ESD damage before you receive them. If actions are taken after components are received, the components are susceptible to receiving ESD damage.

 

1. Improve Grounding

Inside an EPA, all conductors (including people) are grounded. Now you’re probably thinking: “But I’ve already grounded my operators and worksurfaces. What else is there left to do?”. Firstly, well done for properly grounding the ‘objects’ in your EPA. The next step is to adjust and improve your current program to allow for even better protection. Here are some suggestions:

Personnel:

  • Decrease the wrist strap and ESD footwear upper limit. The ESD Association has test data showing charge on a person is less as the path-to-ground resistance is less.
  • Use continuous monitors and ESD smocks
  • Introduce/increase the use of ESD flooring
  • Use sole or full coverage foot grounders (rather than heel grounders)

770750-Use.jpg
Full coverage foot grounders are recommended when handling ultra-sensitive devices

Worksurfaces:

  • Reduce the required limit for Point-to-Point resistance of 1 x 109 per the ESD Standard to 106 to 108 ohms (see #5). The reason for this reduction is simple: 1 x 109 is too high as it still produces thousands of volts of in electrostatic charges. However, the resistance cannot be too small either as this can lead to a sudden ‘hard discharge’ potentially damaging ESD sensitive components.

Other:

  • Improve grounding of carts, shelves and equipment to Ground
  • Minimize isolated conductors like devices on PCBs

 

2. Minimize Charge Generation

The best form of control is to minimize charge generation. First, you should always use shielding packing products like bags or containers (especially when outside an EPA) as these protect from generating charges in the first place. For more information on choosing the correct type of ESD Packaging, we recommend reading this post.

The next step is to eliminate charges once they are generated – this can be achieved through grounding and ionization. We’ll cover ionization in #3 and #4. We’ve already talked about improved grounding in #1. However, for ultra-sensitive components, we also recommend the following:

  • Personnel: Use low-charging floor finish
  • Surfaces: Use low-charging topical antistatic treatments

Both types of ESD products create a low tribocharging coating which allows charges to drain off when grounded. The antistatic properties will reduce triboelectric voltage to under 200 volts.

 

3. Remove Insulators

When talking about conductors and insulators, we explained that insulators cannot be grounded and can damage nearby sensitive devices with a sudden uncontrolled discharge. It is therefore critical to eliminate ALL insulators that are not required in your EPA: plastic cups, non-ESD brushes, tapes etc. How? Here are a couple of options:

  • Replace regular production supplies and fixtures with dissipative, low charging versions, e.g. ESD dissipative brushes, ESD dispensers, ESD tape, ESD Chairs etc.
  • Shield charges on clothing by using ESD smocks.

Workstation.png
Use ESD safe accessories whenever possible

If an insulator is absolutely necessary for production and cannot be removed from the EPA, you could consider a topical treatment which will reduce triboelectric charges.

Is this not an option, then move on to tip #4.

 

4. Use Ionization

First, ionization is not a cure-all. We’ve learnt that ionizers neutralize charges on an insulator.

However, that does not mean that you can just have any insulator in your EPA because the ionizer will “just fix it”. No, in this instance, prevention is generally a better option than the cure. So, your priority should ALWAYS be to remove non-process essential insulators from your EPA – see tip #3. If this is not possible – then ionization becomes essential.

Ionization:

  • Ionizers can be critical to reduce induction charging caused by process necessary insulators
  • Ionizers can be critical in eliminating charges on isolated conductors like devices on PCBs
  • Offset voltage (balance) and discharge times are critical considerations depending on the actual application
  • Ionization can reduce ElectroStatic Attraction (ESA) and charged particles clinging and contaminating products.

It is recommended to use ionizers with feedback mechanisms, so you’re notified if the offset voltage is out of balance.

 

5. Increase ESD Training and Awareness

ESD Training is a requirement of every ESD Program. When handling ultra-sensitive devices, it is even more important to remind everyone what pre-cautions are necessary to avoid damage. Regular ‘refreshers’ are a must and it is recommended to verify the effectiveness of the training program, e.g. through tests. So, who, when and what should be taught?

AdobeStock_105568884.jpeg
ESD Training is a vital part of every successful ESD Control Program

  • ESD training needs to be provided to everyone who handles ESD sensitive devices – that includes managers, supervisors, subcontractors, visitors, cleaners and even temporary personnel.
  • Training must be given at the beginning of employment (BEFORE getting anywhere near a sensitive products) and in regular intervals thereafter.
  • Training should be conducted on proper compliance verification procedures and on the proper use of equipment used for verification.

 

6. Create an enhanced Compliance Verification Plan

We talked in a previous post about compliance verification, what it is and how to create a plan that complies with the ESD standard. So, if you already followed our steps and have a plan in place, here are a few tips to improve your compliance verification plan:

  • Use a computer data collection system for wrist straps and foot grounders testing
  • Increase the test frequency of personnel grounding devices from once per day to every time the operator enters the EPA
  • Use continuous monitors where operators are grounded via wrist straps. Consider computer based monitor data collection system, e.g. SMP. This should include continuous monitoring of the mat Ground.
  • Use Ground continuous monitors, e.g. Ground Master. At a large facility, the most frequent reoccurring violation is the ESD mat ground cord either becoming disconnected from the mat or grounding point. As Ground continuous monitors will only test the fact that the mat is grounded, it is still imperative that the Resistance to Ground of the mat is regularly tested. Remember that the use of improper mat cleaners can raise the mat surface resistance above the upper recommended level of <109
  • Test ionizers more frequently or consider self-monitoring ionizers. Computer based data collection systems are a good alternative, too.
  • Increase the use of a static field meter and nano coulomb testing to verify that automated processes (like auto insertion, tape and reel, etc.) are not generating charges above acceptable limits.

 

Conclusion

“Class 0” items require additional measures of ESD protection due to their sensitivity to ESD damage. The best way to protect these ultra-sensitive components is to increase ESD protective redundancies and periodic verifications to all ESD Control technical elements.

To decrease the probability of ESD damage while handling ultra-sensitive items, additional precautions are required. This includes additional and/or more stringent technical requirements for ESD control products, increasing redundancies, and more frequent periodic verifications or audits.

Additionally, ESD control process systems should be evaluated as to their performance as a system. It is important to understand how the technical elements in use perform relative to the sensitivity of the devices being handled. Thus, tailoring the process to handle the more sensitive parts. For example: If the footwear/flooring allows a person’s body voltage to reach 80 volts and a 50 withstand voltage item gets introduced into the process, you must either allow only handling via wrist straps or would have to find a way to modify the footwear/flooring performance to get peak voltages below the 50 volt threshold.

Remember: The ESD Standard gives recommendations that will always be behind current/future developments. As soon as a Standard is published, technology will have progressed. In order to protect your devices and company reputation for reliable devices – it is recommended your company take responsibility to implement methods/procedures that exceed the recommendations of the ESD Standard to fit your sensitive component requirements.

 

References:

The Importance of Personnel Grounding Testers

We already know that in an ESD Protected Area (EPA) all surfaces, objects, people and ESD Sensitive Devices (ESDs) are kept at the same potential which is achieved by using ‘groundable’ materials that are then linked to ground. We have also learnt that the most common personnel grounding device to link people to ground are wrist straps. People who are moving around should instead wear ESD footwear.

So how do you know if your wrist straps and ESD footwear are working properly? Excellent question! And one we’ll answer in today’s post so let’s jump right in!

Purpose of Personnel Grounding Testers

Wrist straps and ESD footwear should be part of your Verification Plan. Each component in an EPA plays a vital part in the fight against electrostatic discharge (ESD). If just one component is not performing correctly, ESD sensitive devices can be damaged, potentially costing your company thousands of dollars.

Wrist straps and ESD footwear can fail and damage cannot always be detected by visual inspection. Just by looking at the items you would not know if they still provide sufficient protection. Personnel grounding testers should be used to provide feedback to verify the functionality of an operator’s wrist strap and/or footwear.

Your Personnel Grounding Checklist - Wear, Verify, Log, Handle
Your Personnel Grounding Checklist

Your Personnel Grounding Checklist:

  1. Wear your personnel grounding equipment such a wrist strap and/or footwear
  2. Verify your personnel grounding system using a wrist strap and/or footwear tester. Wrist straps and footwear, need to be tested at least daily before handling ESD sensitive devices and should be worn while checking.
  3. Log a record of each test. Records should be kept for quality control purposes.
  4. Handle ESD sensitive components ONLY if your wrist strap and/or footwear pass(es) the test.

Types of Personnel Grounding Testers

Personnel grounding testers can be purchased in two configurations:

  • Wrist strap tester
  • Wrist strap and footwear tester

As wrist straps are the most commonly used personnel grounding device to ground operators, you will find a lot of testers on the market that check wrist straps only. Combined wrist strap and footwear testers will verify your wrist straps AND footwear.

In addition to WHAT the testers verify, you will also be faced with a wide range of devices differing in HOW they test. Below you will see a (by no means complete) list of options:

  • Continuous and split footplate: You will find testers with a continuous footplate which require each foot to be tested separately one after the other. Dual-footplate or independent footwear testers feature a split footplate which allows the unit to verify both feet independently at the same time. This can be an efficient time-saver if you have a number of operators in your company who are required to check their personnel grounding devices.
  • Portable, wall-mountable and fitted testers: Portable battery-powered (predominantly) wrist strap testers are suitable for small labs or for supervisors to spot-check workers and ensure compliance. Wall-mountable units are generally supplied with a wall plate which attaches to a wall; the tester is then mounted on to the wall plate. Some personal grounding devices are accompanied by a stand (and built-in footplate) which allow for a more freely positioning of the unit within a room.
  • Relay terminal: A few testers on the market are fitted with a relay terminal (electrically operated switch) that can be integrated with electronic door locks, turnstiles, lights, buzzers, etc. This can be of advantage if companies only want to allow personnel in an EPA that have passed their wrist strap and/or footwear test.
  • Data acquisition: A growing number of personnel grounding devices allow for test activity data to be logged in a database. The units link to a computer which records operator identification, test results, resistance measurements, time and more. Paperless data can enhance operator accountability, immediately identifying problems while reducing manual logging and auditing costs.

Operation of Personnel Grounding Testers

Wrist strap testing:

If you are not using a continuous or a constant monitor, a wrist strap should be tested at least daily. This quick check can determine that no break in the path-to-ground has occurred. Wrist straps should be worn while they are tested. This provides the best way to test all three components:

  • the wrist band
  • the ground cord (including the resistor)
  • the interface (contact) with the operator’s skin

The SCS Combo Wrist Strap/ Footwear Tester
The SCS Combo Tester can be used to test wrist straps – more information

To ensure that the resistance to ground of personnel is within specification it is important to measure the entire system (i.e., wrist strap, person, and ground connection). The wrist strap system test method is described in ANIS/ESD S1.1. In general, the test method measures the resistance of the ground cord, wristband or cuff, and the interface of the band or cuff of the wearer.” [Handbook ESD TR20.20 Clause 8.2 Wrist Strap System]

The wrist strap system should be tested daily to ensure proper electrical resistance. Nominally, the upper resistance reading should be < 35 megohms or a user-defined resistance. Daily test records can provide evidence of conformity. Daily testing may be omitted if continuous monitors are used.” [ANSI/ESD S1.1 Clause A3. Frequency of System Testing]

If the wrist strap tester outputs a FAIL test result, stop working and test the wrist band and cord individually to find out which item is damaged. Replace the bad component and repeat the test. Obtain a PASS test result before beginning work. For more information on troubleshooting failed wrist straps, check this post.

Footwear testing:

If you are using a flooring / footwear system as an alternative for standing or mobile workers, ESD footwear should be tested independently at least daily while being worn. Proper testing of foot grounders involves the verification of:

  • the individual foot grounder
  • the contact strip
  • the interface between the contact strip and the operator’s perspiration layer

a) Place the foot grounders on the user’s shoes per the manufacturer’s instructions.
b) Place the left foot on the floor plate and touch the body contact area on the tester with one hand. Activate the tester per the manufacturer’s instructions.
c) Remove the left foot from the floor plate.
d) Repeat steps b and c with the right foot.
[ANSI/ESD SP9.2 Clause 6.2.2 Procedure (Integrated Tester)]

The SCS Dual Combination Tester is used to test wrist straps and footwear
The SCS Dual Combination Tester is used to test wrist straps and footwear – more information

If the footwear tester outputs a FAIL test result, stop working, and test the foot grounder and contact strip individually to find out which item is damaged. Replace the foot grounder. Obtain a PASS test result before beginning work.

Conclusion

Wrist straps and footwear need to be tested at least daily before handling any ESD sensitive devices. Personnel grounding devices need to be worn for verification using a wrist strap and/or footwear tester.

A record of each test has to be kept for quality control purposes.

Only handle ESD sensitive components if your wrist strap and/or footwear pass(es) the test.

 

Controlling ESD/EOS During the Soldering Process

When the tip of a soldering iron comes into direct electrical contact with the pins of a sensitive component, there is a danger of voltage and/or current signal transfer between:

  • the grounded iron tip and the grounded PC board,
  • the ungrounded iron tip and the grounded PC board,
  • the grounded iron tip and the ungrounded PC board.

This can cause Electrical Overstress (EOS) and Electrostatic Discharge (ESD).

What is Electrical Overstress (EOS) and why is it important to detect?

EOS is the exposure of a component or PCB board to a current and/or voltage outside its operational range. This absolute maximum rating (AMR) differs from one device to the next and needs to be provided by the manufacturer of each component used during the soldering process. EOS can cause damage, malfunction or accelerated aging in sensitive devices.

ESD can be generated if a component and a board have different potentials and the voltage transfers from one to the other. When such an event happens, the component goes through EOS. ESD can influence EOS, but EOS can also be influenced by other signals.

Many people are familiar with Electrostatic Discharge (ESD) which is caused by the spontaneous discharge between two materials that are at different levels of ElectroStatic potential. Once electrostatic potential between the two materials is balanced, the ESD event will stop.

An EOS event on the other hand is created by voltage and/or current spikes when operating equipment; it can therefore last “as long as the originating signal exists”. [Source] The potentially never-ending stimulus of EOS is what makes it such a big concern in the electronics industry. Even though the voltage levels are generally much lower compared to an ESD event, applying this smaller voltage combined with a larger peak current over a long period of time will cause significant damage.

The high temperatures during an EOS event (created by the high current) can lead to visible EOS damage.

For more information on EOS and the differences to ESD, check-out this post.

Sources of EOS during the Soldering Process

When soldering components, it’s the tip of the soldering iron that comes into contact with the potentially sensitive device. Therefore, many people assume the soldering tip is the cause of ESD/EOS. However, the soldering iron and its tip are just some of the components used at a workbench. Other components on the bench like tweezers, wiring, test equipment, etc. can also be sources of ESD/EOS as they come into contact with the component or board.

There are many sources of EOS during the soldering process, which can include:

  • Loss of Ground
    The tip of an ungrounded soldering iron can accumulate a voltage of up to ½ of the iron’s supply voltage. It can be caused within the soldering iron itself or in power outlets.
  • Noise on Ground
    If a noise signal exists on ground, the tip of the solder iron will carry noise, too. These high-frequency signals, or electromagnetic interference (EMI), are disturbances that affect an electrical circuit, due to either electromagnetic induction or electromagnetic radiation emitted from an external source.
  • Noise on Power Lines
    Noise not only generates via ground but in power lines, too. Transformers and power supplies that convert voltages to 24V are the main culprit. They regularly carry high-frequency spikes which end up on the tip of the soldering iron.
  • Power Tools
    Although not technically related to the soldering process itself, it’s worth mentioning that the tips of power tools (e.g. electric screwdrivers) may not be properly grounded during rotation. This can result in high voltage on the tip itself.
  • Missing/Inadequate ESD Protection
    ESD can be a cause of EOS damage. Therefore, it is essential to have proper ESD Protection in place. A voltage on the operator or the PCB board can otherwise lead to an ESD Event and expose the components on the PCB to EOS.

Detecting EOS during the Soldering Process

EOS/ESD events can be detected, measured, and monitored during the soldering process using a variety of diagnostic tools.

Diagnostic Tools

  • SCS CTM051 Ground Pro Meter
    The SCS CTM051 Ground Pro Meter is a comprehensive instrument that measures ground impedance, AC and DC voltage on the ground as well as the presence of high-frequency noise or electromagnetic interference (EMI) voltage on the ground. It will alert if the soldering iron tip has lost its ground or has EMI voltage induced into the tip from an internal source on the soldering iron or from an EMI noisy ground or power lines.

    CTM051
    The SCS CTM051 Ground Pro Meter
  • SCS CTM048 EM Eye – ESD Event Meter
    The SCS CTM048 EM Eye – ESD Event Meter paired with the SCS CTC028 EM Field Sensor is a diagnostic tool for the detection and analysis of ESD events and electromagnetic fields and can identify sources of harmful ESD Events and electromagnetic interference (EMI).

    CTM048-21
    The SCS CTM048 EM Eye – ESD Event Meter paired with the SCS CTC028 EM Field Sensor

EOS Continuous Monitors

  • SCS CTC331-WW Iron Man® Plus Workstation Monitor
    The SCS CTC331-WW Iron Man® Plus Workstation Monitor is a single workstation continuous monitor which continuously monitors the path-to-ground integrity of an operator and conductive/dissipative worksurface and meets ANSI/ESD S20.20.The Iron Man® Plus Workstation Monitor is an essential tool when it comes to EOS detection. The unit is capable of detecting EOS on boards and alarms if an overvoltage (±5V or less) from a tool such as a soldering iron or electric screwdriver is applied to a circuit board under assembly.

    CTC331-WW
    The SCS CTC331-WW Iron Man® Plus Workstation Monitor

Data Acquisition

  • SCS Static Management Program
    SCS Static Management Program (SMP) continuously monitors the ESD parameters throughout all stages of manufacturing. It captures data from SCS workstation monitors, ground integrity monitors for equipment, ESD event and static voltage continuous monitors and provides real-time data of manufacturing processes.The SCS 770063 EM Aware Monitor, which is part of SMP, can help during the soldering process by monitoring ESD events and change of static voltage that may result in EOS. The EM Aware alarms (visual and audibly) locally and sends data to the database of the SMP system if any of the ESD parameters are detected to be higher than user-defined limits.

    770063.jpg
    The SCS 770063 EM Aware Monitor

Eliminating EOS during the Soldering Process

Once the source of ESD/EOS is known, there are many things that can be done to prevent it in the first place: 

1. Managing Voltage on a PCB board

PCB boards contain isolated conductors and non-conductive (insulative) components. The only way to handle voltage on a PCB board is neutralizing potential static charges through ionization. An ionizer creates great numbers of positively and negatively charged ions. Fans help the generated ions flow over the work area to neutralize static charges (or voltage) on a PCB board in a matter of seconds.

For more information on ionization and how to choose the right type of ionizer for your application, please read these posts.

2. Managing Voltage on an Operator

Static voltage on an operator can be eliminated through proper grounding using a workstation monitor, e.g. WS Aware or Iron Man Plus Monitor, and proper grounding hardware. Sitting personnel are required to wear wrist straps. A wrist strap consists of a conductive wristband which provides an electrical connection to skin of an operator, and a coil cord, which is connected to a known ground point at a workbench, a tool or a continuous monitor. While a wrist strap does not prevent generation of voltages, its purpose is to dissipate these voltages to ground as quickly as possible.

Sitting personnel can also use continuous monitors – not only is the operator grounded through the continuous monitor, but they also provides a number of additional advantages:

  • Immediate feedback should a wrist strap fail
  • Monitoring of operators and work stations
  • Detection of split-second failures
  • Elimination of periodic testing

This post provides more details on continuous monitors.

Moving or standing personnel are grounded via a flooring/footwear system. ESD Footwear (e.g. foot grounders) are designed to reliably contact grounded ESD flooring and provide a continuous path-to-ground by removing electrostatic voltages from personnel.

3. Managing Current

One solution is the “re-routing of ground connection and separation of “noisy” ground from a clean one” as “connecting soldering iron and the workbench to the “quiet” ground often result in lower level of transient signals.“. [Source]

This will greatly reduce the high-frequency noise that could cause EOS damage.

If the noise on power lines and ground cannot be reduced manually, then the use of noise filters becomes necessary to reduce the risk of EOS exposure during the soldering process. Utilizing these filters suppresses the noise on power lines and will allow the solder iron to use “clean” power only.

In his papers, Vladimir Kraz, explains the set-up of a soldering station using a noise filter in more detail.

Noise-Filter
Soldering Iron with Power Line EMI Filter [Source]

Conclusion

During the soldering process, current and voltage spikes between the solder tip and PCB can cause ESD/EOS. Sources are varied and can include:

  • Loss of Ground
  • Noise on Ground
  • Noise on Power Lines
  • Power Tools
  • Missing/Inadequate ESD Protection

ESD/EOS can be identified and controlled using diagnostic tools. SCS offers a number of tools that can detect current, voltage and EMI – all potentially leading to ESD and EOS.

Once the source of ESD/EOS is known, the next step is eliminating the source:

  • Managing voltage on a PCB board using ionizers.
  • Managing voltage on an operator using workstation monitors or foot grounders.
  • Managing current using noise filters.
  • Managing voltage on materials at the work bench.
  • Managing ESD generation during specific processes.
  • Managing grounding.

 For more information regarding this topic, please see below for additional references.

References:

Checking your ESD Control Products

Last time we explained how to easily create a compliance verification plan and why it’s important to have one in place. Today’s post will elaborate on the subject of periodic verification and highlight common products in your EPA that should be regularly verified and more importantly how they should be checked.

Why periodic verification
Compliance verification is a requirement of ANSI/ESD S20.20:
The Organization shall prepare an ESD Control Program Plan that addresses each of the requirements of the Program. Those requirements include:
– Training
– Product Qualification
– Compliance Verification
– Grounding / Equipotential Bonding Systems
– Personnel Grounding
– ESD Protected Area (EPA) Requirements
– Packaging Systems
– Marking” [ANSI/ESD S20.20 clause 7.1 ESD Control Program Plan]

Installed ESD Control products must be checked regularly to ensure they meet the required limits per the ESD Standard. “Compliance verification records shall be established and maintained to provide evidence of conformity to the technical requirements. The test equipment selected shall be capable of making the measurements defined in the Compliance Verification Plan.” [ANSI/ESD S20.20 clause 7.3 Compliance Verification Plan]

Below, you will find a list of the most common ESD Control Products in your EPA and how to test them:

Worksurface Matting
The purpose of ESD workbench matting is to ensure that when charged conductors (conductive or dissipative) are placed upon the surface, a controlled discharge occurs and electrostatic charges are removed to ground. However, this only occurs if the ESD worksurface is connected to ground. If the matting is out-of-spec, not grounded at all, or the stud on the mat has become loose or if the ground cord has become disconnected, charges cannot be removed.
Many companies use a daily checklist, which includes the operator having to verify that ground cords are firmly connected.
Remember to regularly clean your workbench matting to maintain proper electrical function. Do not use cleaners with silicone as silicone build-up will create an insulative film on the surface.
The company’s compliance verification plan should include periodic checks of worksurfaces measuring:

  • Resistance Point-to-Point (Rp-p) and
  • Resistance-to-ground (Rg)

Testing a worksurface using SRMETER2

A surface resistance tester can be used to perform these tests in accordance with ANSI/ESD S20.20 and its test method ESD TR53; if these measurements are within acceptable ranges, the worksurface matting and its connections are good.

Wrist Straps
As discharges from people handling ESD sensitive devices cause significant ESD damage, the wrist strap is considered the first line of ESD control.
Before handling ESD sensitive items, you should visually inspect the wrist strap to see if there are any breakages etc. The wrist strap should then be tested while worn using a wrist strap tester. This ensures all three components are checked: the wrist band, the ground cord (including resistor) and the contact with the operator’s skin. Records of each test should be kept. Wiggling the resistor strain relief portion of the coil cord during the test will help identify failures sooner. Analysis and corrective action should take place when a wrist strap tester indicates a failure.

Checking wrist straps using 746

It is recommended that wrist straps are checked at least daily. An even better solution to daily wrist strap checks is the use of continuous monitors. They will alarm if the person or worksurface is not properly grounded.

A note on worksurface matting and wrist straps: if you are using common ground points to ground the operator and/or work surface matting, remember to measure resistance to ground regularly as well (every 6 months for example).

Floor Matting
A flooring / footwear system is an alternative for personnel grounding for standing or mobile workers. Foot grounders quickly and effectively drain the static charges which collect on personnel during normal, everyday activities. Foot grounders should be used in conjunction with floor surfaces which have a surface resistance of less than 1010 ohms.
As ESD floors get dirty, their resistance increases. For optimum electrical performance, floor matting must be cleaned regularly using an ESD mat cleaner. Do not use cleaners with silicone as silicone build-up will create an insulative film on the surface.
Dissipative floor finish can be used to reduce floor resistance. Periodic verification will identify how often the floor finish needs to be applied. As the layer(s) of dissipative floor finish wear, the resistance measurements will increase. So, after some amount of data collection, a cost-effective maintenance schedule can be established.
Floor matting can be checked using a resistance meter. A surface resistance meter is designed to measure resistance point-to-point (Rp-p) or surface to ground (Rg) in accordance with ANSI/ESD S20.20 and its test method ESD TR53.

Footwear
ESD Shoes or foot grounders play an essential part in the flooring/footwear system.
Before handling ESD sensitive devices, visually inspect your ESD footwear for any damage. Just like wrist straps, footwear should be checked while being worn using a wrist strap/footwear tester.

Checking foot grounders using 770750

Records of each test should be kept. Analysis and corrective action should take place when a footwear tester indicates a failure. Footwear needs to be checked daily.

ESD Packaging
Re-using shielding bags is acceptable as long as there is no damage to the shielding layer. Shielding bags with holes, tears or excessive wrinkles should be discarded.

An operator packing an ESD sensitive item into a Shielding Bag
Make sure your ESD shielding bags are un-damaged

It is up to the user to determine if a shielding bag is suitable for re-use or not. The testing of every bag before re-use is not practical. Many companies will discard the shielding bag once used and replace it with a new one. Others will use a system of labels to identify when the bag has gone through five handling cycles:

  • Non-reusable labels are used that require the label be broken to open the bag.
  • The bag is then resealed with a new label.
  • When there are five broken labels, the bag is discarded.

The same principle applies to other ESD packaging, e.g. component shippers.

Ionizers
Ionizers are intended to neutralize static charges on insulators thereby reducing their potential to cause ESD damage. However, poorly maintained ionizers with dirty emitter pins and out-of-balance ionisers can put a charge on ungrounded items.
Remember to clean ionizer emitter pins and filters regularly. You can now even purchase ionizers that will alarm when emitter pins need to be cleaned or the ionizer is out of balance.

Charge plate monitor and static decay measurements using 963E ionized air blower

Static neutralization (the ability to reduce or eliminate a charge on a surface) is an important quality for ionizers. Static decay time is defined as the time interval needed to reduce a defined voltage potential on an object to a defined lower potential by means of applied ionized air. Another important aspect for ionizers is the ability to produce a balanced stream of positive and negative ions. A charged plate monitor or equivalent can be used to accurately measure both of these parameters.
For more detailed information on measuring the performance of ionizers refer to the ESD standard ANSI/EOS/ESD-S3.1 for Protection of Electrostatic Discharge Susceptible Items-Ionization.

Wrist Strap/Footwear and Resistance Testers etc.
So, you check your wrist straps and/or footwear and workbench and/or floor matting regularly. But have you remembered the testers themselves? What good do all the checks do, if the testers you use are out-of-spec and show you incorrect results?
Yearly calibration is recommended – many manufacturers offer a calibration service or alternatively you can purchase calibration units from them and perform the calibration yourself.

There you have it – a list of the most commonly used products in your ESD Protected Area (EPA) that you should check on a regular basis.
Questions for you: Do you have a verification plan in place? If so, how often do you check your ESD protection products?

The importance of an ESD Protected Area (EPA)

In our last post, we talked about ESD: what it is, what types of ESD damage there are and what costly effects ESD can have. Missed our very first post? Catch-up here.
All caught up? Right, moving on. Today you will learn how to avoid ESD damage and protect your ESD sensitive items. So, let’s jump right in.

The fundamental ESD Control Principles
We’ve established that ESD is the hidden enemy in the electronics industry. Therefore, the BIG question is: how exactly do you control ElectroStatic Discharge (ESD) in your workplace? Easy – just follow these ESD fundamentals:

  1. Ground all conductors including people
  2. Remove all unnecessary non-conductors (also known as insulators)
  3. Place ESD sensitive devices inside of shielding packaging when transported outside of an ESD Protected Area (EPA)

Per ESD Handbook ESD TR20.20-2008 section 2.4 “It should be understood that any object, item, material or person could be a source of static electricity in the work environment. Removal of unnecessary nonconductors, replacing nonconductive materials with dissipative or conductive materials and grounding all conductors are the principle methods of controlling static electricity in the workplace, regardless of the activity.

These are the essential principles of ESD Control. If you implement all three points above, you will be in control of ESD and your sensitive items will be protected. Well, that wasn’t hard, was it? Don’t be terrified – we’ll go through everything in detail. We’ll cover #2 and #3 in future points – today’s focus is #1.

Definition of an ESD Protected Area (EPA)
An ESD Protected Area (EPA) is a designated zone – all surfaces, objects, people and ESD Sensitive Devices (ESDs) within are kept at the same electrical potential. This is achieved by simply using ‘groundable’ materials for covering of surfaces and for the manufacture of containers and tools. This applies to all items with an electrical resistance of less than 109 ohms.

An EPA could be just one workstation or it could be a room containing several different workstations. It can be portable as used in a field service situation or permanent.

Example-EPA-Area
Example of an ESD Protected Area

The user guide CLC/TR 61340-5-2:2008 defines an EPA as follows:
An ESD protected area (EPA) is an area that is equipped with the ESD control items required to minimize the chance of damaging ESD sensitive devices. In the broad sense, a protected area is capable of controlling static electricity on all items that enter that work area. Personnel and other conductive or dissipative items shall be electrically bonded together and connected to ground (or a common connection point when a ground is not available) to equalize electrical potential among the items. The size of an EPA can vary greatly. A protected area may be a permanent workstation within a room or an entire factory floor encompassing thousands of workstations. A protected area may also be a portable worksurface or mat used in a field service situation.” [CLC/TR 61340-5-2:2008 Use guide clause 4.6 Protected areas (EPA)]

You’re probably wondering now, how exactly you can get all surfaces, objects and operators to the same electrical potential. Fear not – we’ve got you covered!

  1. Personnel Grounding
    As previously stated, a fundamental principle of ESD control is to ground conductors including people at ESD protected workstations.Wrist straps are the first line of defense against ESD, the most common personnel grounding device used, and are required to be used if the operator is sitting. The wristband should be worn snug to the skin with its coil cord connected to a common point ground which is connected to ground, preferably equipment ground.

    Wearing-Wrist-Strap
    Wearing a wrist strap and connecting it to a common point ground

    If you are not using a continuous or a constant monitor, a wrist strap should be tested while being worn at least daily. This quick check can determine that no break in the path-to-ground has occurred. Part of the path-to-ground is the perspiration layer on the person; an operator with dry skin may inhibit the removal of static charges and may cause a test failure.
    The wrist strap system should be tested daily to ensure proper electrical value. Nominally, the upper resistance reading should be ” [ANSI/ESD S1.1 Annex A, 3 Frequency of Functional Testing]

    A Flooring / Footwear system is an alternative for personnel grounding for standing or mobile workers. Foot grounders or other types of ESD footwear are worn while standing or walking on an ESD floor. ESD footwear is to be worn on both feet and should be tested independently at least daily while being worn. Unless the tester has a split footplate, each foot should be tested independently, typically with the other foot raised in the air.
    Compliance verification should be performed prior to each use (daily, shift change, etc.). The accumulation of insulative materials may increase the foot grounder system resistance. If foot grounders are worn outside the ESD protected area testing for functionality before reentry to the ESD protected area should be considered.” [ESD SP9.2 APPENDIX B – Foot Grounder Usage Guidance]

    Both ESD footwear and ESD floor are required. Wearing ESD footwear on a regular, insulative floor is a waste of time and money.

    Wearing-Foot-Grounders
    Wearing foot grounders on an ESD floor

    Part of the path-to-ground is the perspiration in the person’s shoes. The conductive tab or ribbon of foot grounders should be placed inside the shoe under the foot with the excess length tucked into the shoe. Thanks to the perspiration in the shoe, direct contact with the skin is normally not necessary.

    If an operator leaves the EPA and walks outside wearing ESD footwear, care should be taken not to get the ESD footwear soiled. Dirt is typically insulative, and the best practice is to re-test the ESD footwear while being worn each time when re-entering the EPA.

  2. Working Surfaces
    ESD working surfaces, such as mats, are typically an integral part of the ESD workstation, particularly in areas where hand assembly occurs. The purpose of the ESD working surface is two-fold:

    1. To provide a surface with little to no charge on it.
    2. To provide a surface that will remove ElectroStatic charges from conductors including ESDS devices and assemblies) that are placed on the surface.

    ESD mats need to be grounded. A ground wire from the mat should connect to the common point ground which is connected to ground, preferably equipment ground. For electronics manufacturing a working surface resistance to ground (RG) of 1 x 106 to less than 1 x 109 ohms is recommended.
    The single most important concept in the field of static control is grounding. Attaching all electrically conductive and dissipative items in the workplace to ground allows built-up electrostatic charges to equalize with ground potential. A grounded conductor cannot hold a static charge.” [Grounding ANSI/ESD S6.1 Foreword]
    Per ANSI/ESD S20.20 section 6.2.1.2 Grounding / Bonding Systems Guidance, “In most cases, the third wire (green) AC equipment ground is the preferred choice for ground.
    Best practice is that ground connections use firm fitting connecting devices such as metallic crimps, snaps and banana plugs to connect to designated ground points. The use of alligator clips is not recommended.

    The working surface must be maintained and should be cleaned with an ESD cleaner. Regular cleaners typically contain silicone, and should never be used on an ESD working surface. ESD Handbook ESD TR20.20-2008 section 5.3.1.14 Maintenance “Periodic cleaning, following the manufacturerís recommendations, is required to maintain proper electrical function of all worksurfaces. Ensure that cleaners that are used do not leave an electrically insulative residue common with some household cleaners that contain silicone.

  3. Other moveable objects
    Moveable items (such as containers and tools) are grounded when placed on a grounded surface or being held by a grounded operator. Everything that does not readily dissipate charge must be excluded from the EPA (refer to #2 of our ESD Control Principles above). Regular plastics, polystyrene foam drink cups and packaging materials, etc. are typically high charging and have no place at an ESD protective workstation.

    Intention of an ESD Protected Area (EPA)We’ve learnt in our previous blog post that ElectroStatic discharge (ESD) can damage components and products that contain electronics. A lot of the time, this damage is not detected during quality inspection and can cause significant problems further down the line.An ESD Protected Area (EPA) is an area that has specifically been created to control ESD; its purpose is therefore to avoid ALL problems resulting from ESD damage. Workers need to understand AND follow the basics of ESD control to limit the generation of electrostatic charges as well as limit and slow discharges in the EPA.Recognizing an ESD Protected Area (EPA)
    An ESD Protected Area must be clearly identified using signs and/or aisle tape. This ensures operators and visitors are alerted when entering (or leaving) an ESD Protected Area which require special precautions (grounding via wrist straps and/or foot grounders etc.). It also indicates that they are entering (or exiting) areas where exposed ESDS items can be handled safely.Remember to be consistent throughout your shop floor, i.e. use the same signs. This will avoid confusion for your operators.

    EPA-Caution-Sign
    Example of an EPA caution sign

    While signs are one way of indicating the boundaries of an EPA, it is not the only way. Any alternate method that alert the personnel that an EPA begins is acceptable to ANSI/ESD S20.20. Some of the alternate ways to mark the boundaries of an EPA are:

    • tape on the floor
    • different color floor tiles
    • different color carpet
    • any other way to establish boundary conditions

    Anyway to distinguish the boundaries of an EPA would be acceptable as long as the personnel are aware of the indications and take the proper precautions while inside the EPA.” [ESD TR20.20-2016 section 9.1.2 EPA Boundary Indicators]

    Building an ESD Protected Area (EPA)
    A basic form of an ESD Protected Area is a workstation consisting of the following components:

    • An ESD working surface mat
    • A grounding cord
    • A wristband
    • A coiled cord
    • A common point ground

    To set-up an EPA:

    1. Connect the ESD working surface mat to the common point ground using the grounding cord.
    2. Link the operator to the common point ground using the wristband and coiled cord.

    Congratulations – you’ve just created an ESD Protected Area!
    By following the above steps, each component (the ESD mat and the operator) is kept at the same electrical potential (ground). Any ElectroStatic charge (ESD) is removed to ground via the common point ground.